Thermo HILIC Separations Technical Guide 
(請下載檔案)

Thermo HILIC Care Guidelines:
​pH stability: pH range is 2 - 8.

管柱清洗:
Prior to storage, buffered solutions used during HILIC analysis should be washed out of the column (and the HPLC system). A suitable wash solution is 60% organic solvent in water (unless otherwise specificed in product manuals). 
 

管柱保存:
HILIC columns should be stored under HILIC conditions. A mixture of organic solvent and water in a 90:10 ratio is recommended for storage.Ensure the column end-fittings are sealed with end-plugs to prevent the stationary phase from drying.